Skip Navigation

This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (3)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Kahan, E.
Right arrow Articles by Weingarten, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kahan, E.
Right arrow Articles by Weingarten, M. A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Family Practice Vol. 20, No. 6, 730-732
© Oxford University Press 2003, all rights reserved


Article

Comparison of blood pressure measurements on the bare arm, below a rolled-up sleeve, or over a sleeve

Ernesto Kahana,, John Yaphea,b, Hadas Knaani-Levinzb and Michael A. Weingartena,b

a Department of Family Medicine, Sackler School of Medicine, Tel Aviv University, Tel Aviv and b Department of Family Medicine, Rabin Medical Center, Beilinson Campus, Petah Tiqva, Israel

Correspondence to E. Kahan, Head, Epidemiology, Department of Family Medicine, Tel Aviv University, Arazim 4/7, Kfar Saba 44456, Israel; E-mail: ekahan{at}post.tau.ac.il

Received 7 January 2003; Revised 9 June 2003; Accepted 14 July 2003.

Kahan E, Yaphe J, Knaani-Levinz H and Weingarten MA. Comparison of blood pressure measurements on the bare arm, below a rolled-up sleeve, or over a sleeve. Family Practice 2003; 20: 730–732.


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Discussion
 References
 
Background. This study examined the effect of measuring blood pressure below subjects' rolled-up sleeves, over the sleeve, or on the bare arm. This is an important day-to-day issue for the busy GP.

Methods. The sample consisted of 201 subjects in family practice clinics and residents of a senior citizens' home. A digital device was used in all cases. Each participant underwent three blood pressure measurements in each of the following conditions in random order: cuff on bare arm; cuff over the sleeve; and cuff below the rolled-up sleeve. Differences between measurements were plotted against the mean blood pressure. Confounding factors controlled for were age, sex, clothing pressure and skin-fold thickness.

Results. Differences in mean blood pressure readings between the clothed and bare arm were 0.5 mmHg (SD 7.5) for systolic pressure and 1 mmHg (SD 5) for diastolic pressure; neither difference was significant. However, in hypertensive subjects (>140 mmHg systolic), although the mean difference remained small (systolic pressure, 2 mmHg, SD 10), the range of difference for individual subjects was -32 mmHg to +22 mmHg.

Conclusion. The degree of clothing under the sphygmomanometer cuff does not have a clinically important effect on the blood pressure measurement. In patients known or found to be hypertensive, measurement on the bare arm is recommended.

Keywords. Blood pressure, blood pressure cuff, family medicine, measurements, screening.


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Discussion
 References
 
Many factors influence the accuracy of blood pressure measurement, such as inadequate rest, white-coat effects and end-digit preference. Most studies recommend that blood pressure be measured on the bare arm,1–4 although one small study found no significant effect of clothing on blood pressure recordings.5 The present study was undertaken to determine the effect of different methods of blood pressure measurement in ambulatory care.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Discussion
 References
 
Sample
Subjects (n = 201) were recruited from consecutive patients attending a family medicine practice and from residents of a nursing home. All patients were over 15 years old and consented to participate. The sample size was calculated to obtain the number of subjects needed for statistical significance when the estimated proportion of individuals with significant differences between the methods for blood measurements is 50% (‘safest’ choice),6 confidence level of 95% and absolute variation of 10% points (40–60%). The result (n = 96) was then doubled to increase the power in the event of individual variations.

The Association of Advancement of Medical Instrumentation (AAMI) standard for comparison of blood pressure measurement methods7 requires a minimum of 85 patients and at least 225 measurements with each technique. Good agreement is defined as a mean difference of less than 5 mmHg between measurements with a standard deviation of less than 8 mmHg.

Procedure. Subjects rested for 5 minutes before all measurements. They were asked not to smoke, physically exert themselves or eat for 30 minutes before the examination. To avoid measurement and observer biases, the following conditions were set:

  1. only one observer (HK-L), a physician researcher specially trained for this study, performed all the blood pressure measurements;
  2. all measurements were made with the subject seated and his/her arm resting on a smooth surface, with the forearm bent at the level of the heart; and
  3. a digital recording sphygmomanometer (UA-767, A & D Engineering, Inc., Milpitas, CA, USA), equipped with an oscillometric pressure sensor was used in all cases.

Nine measurements were taken consecutively without delays for each subject, three in each of the following three situations, in random order:

With the subject's sleeve rolled above the upper border of the blood pressure cuff (‘rolled sleeve’).

With the cuff placed over the sleeve on the subject's upper arm (‘sleeve’).

With the cuff placed on the subject's bare and unclothed arm (‘bare arm’).

The first of each set of three readings was discarded, and the average of the second and third readings was used for comparison with the readings made in the other two situations.

Other measurements included body mass index (BMI), skin-fold thickness, clothing thickness, and manometric assessment of clothing pressure on the arm.

Statistical analysis and interpretation of data. All data were analysed using analysis of variance (ANOVA) for repeated measures for type of measurement (‘rolled sleeve’, ‘sleeve’ or ‘bare arm’) on blood pressure after controlling for the effects of the confounding factors of age, sex, use of medications, body fat (assessed by BMI and skin-fold thickness), and clothing thickness (for measurements made on the sleeve). Differences between pairs of readings were plotted against the mean of the two readings to assess agreement between the methods of measurement, as described by Bland and Altman.8 Results were considered significant when P < 0.05.

Agreement of readings was compared to the AAMI standard for comparison of blood pressure measurement methods7 and the British Hypertension Society (BHS) protocol for the evaluation of blood pressure measuring devices.9,10 The BHS standard rates devices into four categories (A to D) according to the distribution of the readings made by the different methods within 5, 10 or 15 mmHg of each other. The subjects served as their own controls.

Results
Characteristics of the study population (201 subjects) are shown in Table 1. Mean age was 46 years, 66% were female and mean BMI was 26. Nineteen percent of the patients were obese (BMI > 30), and 17% were receiving antihypertensive medication.


View this table:
[in this window]
[in a new window]
 
TABLE 1 Demographic and anthropometric variables in the study population (n = 201)

 
Table 2 shows differences in blood pressure by method of measurement. For systolic pressure, the mean difference between measurements made on the bare arm compared with a rolled sleeve was -0.54 mmHg (SD 7.4), and for diastolic pressure, 0.56 mmHg (SD 4.6). Corresponding values for ‘bare arm’ versus ‘sleeve’ were 0.02 mmHg (SD 7.6) and 1.27 mmHg (SD 5.6), and for ‘sleeve’ versus ‘rolled sleeve’, 0.51 mmHg (SD 7.5) and -0.7 mmHg (SD 5.8). All findings fell within the AAMI standard for good agreement of blood pressure measurement methods. On the BHS protocol, differences between methods fell into category B for systolic pressure and category A for diastolic pressure, indicating an acceptable level of agreement. The Bland-Altman plots indicated a greater deviation from the mean difference for higher systolic blood pressures (above 140 mmHg) between readings on the bare arm and rolled-up sleeve (Fig. 1). This difference was not apparent for diastolic pressure.


View this table:
[in this window]
[in a new window]
 
TABLE 2 Mean systolic and diastolic blood pressures (mmHg) (and standard deviations) by method of measurement

 


View larger version (11K):
[in this window]
[in a new window]
 
FIGURE 1 Bland-Altman plot of differences in systolic blood pressure readings between bare arm and rolled-up sleeve (mean of two readings)

 
Using ANOVA, the interaction of type of measurement on the variation of systolic and diastolic blood pressure by order of measurement (controlled for co-variants age, body fat, clothing thickness) did not significantly affect the observed differences in blood pressure by method of measurement.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Discussion
 References
 
This study did not demonstrate a significant effect of the degree of clothing (‘rolled sleeve’, ‘sleeve’ or ‘bare arm’) on the results of the blood pressure measurements. This confirms the findings of an earlier small and under-powered study (36 subjects) by Holleman et al.5 The implication of these findings for the busy practitioner is that pedantic concern with technique should not take precedence over performance of the measurement. Our study also showed that in patients with hypertension (systolic blood pressure over 140 mmHg), there may be a greater deviation from the mean difference between clothed and bare-arm readings seen at lower pressures. Therefore, when the blood pressure reading indicates hypertension, or in patients with known hypertension, measurements should be taken under standardized conditions (i.e., on the bare arm), as suggested by Holleman et al.5


    Acknowledgments
 
This study was performed in part fulfilment of the requirements for the MD degree at the Sackler School of Medicine of Tel Aviv University by HK-L.


    References
 Top
 Abstract
 Introduction
 Methods
 Discussion
 References
 
1 O'Brien E, Fitzgerald D, O'Malley K. Blood pressure measurement: current practice and future trends. Br Med J 1985; 290: 729–734.[ISI][Medline]

2 Ataman SL, Cooper R, Rotimi C et al. Standardization of blood pressure measurement in an international comparative study. J Clin Epidemiol 1996; 49: 869–877.[CrossRef][ISI][Medline]

3 Baker RH, Ende J. Confounders of auscultatory blood pressure measurement. J Gen Intern Med 1995; 10: 223–231.[ISI][Medline]

4 Bailey RH, Bauer JH. A review of common errors in the indirect measurement of blood pressure. Sphygmomanometry. Arch Intern Med 1993; 153: 2741–2748.[Abstract]

5 Holleman DR, Westman EC, McCrory DC, Simel DL. The effect of sleeved arms on oscillometric blood pressure measurement. J Gen Intern Med 1993; 8: 325–326.[ISI][Medline]

6 Lwanga SK, Lemeshow S. Sample size determination in Health Studies. A practical manual. Geneva: WHO, 1991.

7 White WB, Berson AS, Robbins C et al. National standard for measurement of resting and ambulatory blood pressure with automated sphygmomanometers. Hypertension 1993; 21: 504–509.[Abstract/Free Full Text]

8 Bland JM, Altman DG. Statistical method for assessing agreement between two methods of clinical measurement. Lancet 1986; i: 307–310.

9 O'Brien E, Coats A, Owens P et al. Use and interpretation of ambulatory blood pressure monitoring recommendations of the British Hypertension Society. Br Med J 2000; 22: 1128–1134.

10 Naschitz JE, Gaitini L, Lowenstein L, Keren D, Tamir A, Yeshurun D. Rapid estimation of the accuracy of automatic blood pressure measuring devices (READ). J Hum Hypertens 1999; 13: 443–447.[CrossRef][ISI][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
CMAJHome page
G. Ma MD, N. Sabin MD, and M. Dawes MBBS MD
A comparison of blood pressure measurement over a sleeved arm versus a bare arm
Can. Med. Assoc. J., February 26, 2008; 178(5): 585 - 589.
[Abstract] [Full Text] [PDF]


Home page
CMAJHome page
D. W. McKay PhD
Measuring blood pressure: A call to bare arms?
Can. Med. Assoc. J., February 26, 2008; 178(5): 591 - 593.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (3)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Kahan, E.
Right arrow Articles by Weingarten, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kahan, E.
Right arrow Articles by Weingarten, M. A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?